Biolight S10 Patient Monitor
The Biolight S10 is a sleek touch screen controlled patient monitor – super ease-of-use and affordability for your department
Perfect for General Practice Clinic or Triage rooms
- 10″ Touch Screen
- NIBP
- SpO2
- 3-Lead ECG (5-Lead optional)
- 2-Temp
- HR/RR
Ex Stock, custom upgrades 4-5 weeks
Description
Biolight S10 Patient Monitor
Medspares is the only offical distributor of Biolight in New Zealand
The BLT S series patient monitor is based on Biolight’s 27 years of experience in the design and manufacture of world-class and innovative medical devices, inspired by the needs of the hospital. Concise and ergonomic product design with new software architecture and cutting-edge technology, S series monitor changes the way the medical staff work and meets the clinical demands.
The Biolight S10 is a compact, touchscreen-controlled patient monitor designed for efficient, real-time monitoring in clinical environments. It features a high-resolution capacitive display with gesture-based operation, offering intuitive usability. The S10 supports key parameters including ECG (3/5/6 lead), SpO₂, NIBP, RESP, PR, and temperature. With a fanless design to reduce cross-contamination risk and up to 8 hours of battery life, it’s ideal for both bedside and mobile use. The monitor includes Early Warning Scoring (EWS), Glasgow Coma Scale (GCS), and optional drip monitoring, enhancing patient safety through early detection and intervention
Main Features BLT-S10:
- Size:168mm×288mm×236mm
- Weight: <4kg
- 10.4-inch Color touch screen
- 800×600 resolution
- Accessory storage box
- EWS
- GCS
Base Specifications:
- 3/5-Lead ECG (3 lead cable included)
- BLT SpO2 – Adult Probe Included
- NIBP – Adult Cuff Included
- HR/PR
- RESP (from pleth)
- 2-temperature (1 probe included)
- 2500mAH Battery
- Printer
Additional information
Weight | 10 kg |
---|
Downloads
Biolight_S12_S10_Milti-monitor-compressed
Biolight_S10&S12_Service_Manual
Biolight_S10&S12_User_Manual
Biolight_P_S Series_printing_through_pc guide-compressed
BLT_P_S_Series_Training
Videos
Specifications
Biolight S10 Patient Monitor – Specifications
NIBP Measurement:
Parameters: SYS, DIA, MAP,PR
Mode of operation: Manual, Auto, STAT, Sequence
Intervals for periodic measurement time:
1min, 2 min, 2.5 min, 3 min, 5 min, 10 min, 15 min, 20 min, 30 min, 1h, 1.5h, 2h, 3h, 4h, 8h
Sequence mode: Up to 5 groups are supported, and each group individually sets
the interval and number of periodic measurement.
STAT mode cycle time: 5 min
Measurement range of cuff pressure: 0~300 mmHg
Initial inflation pressure:
Adult: 100~280mmHg, default 160mmHg
Pediatric: 100~240 mmHg, default 130mmHg
Neonatal: 60~140mmHg, default 100mmHg
Sensor calibration time: One year (recommend)
Unit: mmHg, kPa
Dynamic pressure measurement range:
Systolic
Adult: 30~270 mmHg
Pediatric: 30~235 mmHg
Neonatal: 30~135 mmHg
Diastolic
Adult: 10~220 mmHg
Pediatric: 10~220 mmHg
Neonatal: 10~110 mmHg
Mean
Adult: 20~235 mmHg
Pediatric: 20~225 mmHg
Neonatal: 20~125 mmHg
Dynamic Pressure Measurement Error of Simulator:
Systolic 35~255mmHg: ±8 mmHg
Diastolic 15~195mmHg: ±8 mmHg
Mean 22~215mmHg: ±8 mmHg
Static pressure: ±3 mmHg(±0.4kPa)
Accuracy
Pressure resolution 1 mmHg or 0.1kPa
PR measurement range and accuracy:
Measurement range: 40 ~ 240 bpm
Accuracy ±3bpm or ±3%, whichever is greater
Maximum measurement time:
Neonatal: <90s
Adult, Pediatric: <120s
First overvoltage protection point:
Adult: 297±3 mmHg
Neonatal: 147±3 mmHg
Pediatric: 252±3 mmHg
Second overvoltage protection point:
Adult: 315±10 mmHg
Neonatal: 155±10 mmHg
Pediatric: 265±10 mmHg
Note: The accuracy of NIBP cannot be determined by using a simulator, but under many conditions, it is still necessary to use a simulator to test its performance (for example, a simulator is required for quality control in the production process), and the simulator of the model specified by the manufacturer shall be used for this performance test.
SpO2
Measurement parameter: SpO2, PR, PI and RR, SpO2 waveform and respiration waveform.
Sensitivity: High, Medium, Low
Measurement range and accuracy:
Measurement range 0~100%
Clinical accuracy: 70%~100%: ≤3% (SpO2 probe included in Appendix);
0~69%: unspecified
Simulator measurement error*: 70%~100%: ±2% / 0~69% unspecified
SpO2 Update period: Normal ≤2s
Maximum ≤25s: SpO2 Response time Sensitivity:
High ≤8s
medium ≤11s
Low ≤15s
PR
Measurement range: 25 bpm ~300 bpm
Resolution 1 bpm
Accuracy ± 3bpm
PI
Measurement range 0.05~20.00%
Resolution 0.01%
Accuracy: 0.05~20.00%: ±0.1% or ±10% of reading, whichever is greater
RESP (from pleth)
Measurement range 0 rpm ~90 rpm
Resolution 1 rpm
Accuracy ±2 rpm
Temp
Parameter T1,T2,TD
Probe YSI400 series probe (2252Ω@25℃, accuracy ±0.1ºC)
Measurement site: Surface and coelom
Measurement range and accuracy:
Measurement range: 0.0ºC ~50.0ºC
Resolution 0.1ºC
Accuracy of circuit: ±0.1℃
Power supplied to probe: <20μW
Updated time: Every about 1~2s
Minimum measurement time:
Surface: ≤100s
Coelom: ≤80s
ECG
Lead type: 3-lead: I, II, III (standard)
5-lead: I, II, III, aVR, aVL, aVF, V-(optional lead)
6-lead: I, II, III, aVR, aVL, aVF,Va, Vb (optional lead)
12-lead: I, II, III, aVR, aVL, aVF,V1~V6 (optional module/lead upgrade)
Auto: identify leads automatically
Indication of lead-off shall be provided: Every electrode
ECG abnormal work indications: Every amplification channel shall have an indication of abnormal ECG operation (polarization).
Bandwidth (-3dB):
Diagnostic mode: 0.05~150Hz
Monitor mode: 0.5~40Hz
Operation mode: 1~25Hz
ST mode: 0.05~40Hz
Signal quality display: Expression way: numerical display and waveform color.
Defibrillation Protection:
Breakdown Voltage: 4000V 50Hz/60Hz
Anti-defibrillation effect protection: baseline recovery time: 5s (after defibrillation).
Input signal range:
Input signal range -10.0mV~+10.0mV
|Electrode offset potential ±500 mV d.c.
Input impedance ≥5.0MΩ
System noise ≤30μVpp RTI
Waveform sweep speed:
Display 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s, error ≤±5%
Recorder 12.5mm/s, 25mm/s, 50mm/s, error ≤±5%
Waveform Gain:
Display Å~0.25, Å~0.5, Å~1 (10mm/mV), Å~2, Å~4, error ≤±5%. Auto
Recorder Å~0.25, Å~0.5, Å~1 (10mm/mV), Å~2, Å~4, error ≤±5%.
CMRR:
Diagnostic mode ≥100 dB
Monitor, Operation mode: ≥110 dB
Calibration voltage ≤±5%(Å~1)
|Input offset current <0.1uA
Time constant:
Monitoring mode: ≥0.3s
Diagnostic mode: ≥3.2s
Measurement parameter: HR, PVCs, ST, QT, and Arrhythmia analysis
Multi-lead synchronous analysis function:
Supports synchronous analysis of at least 2 leads, one of which is the key monitoring lead and the other is the auxiliary lead. It is on except 3-lead mode.
Smart Lead Switch:
Automatic, Manual; Default manual (3-lead mode is fixed as manual)
Automatic mode: the algorithm automatically identifies the current smart leads, and the host automatically switches the key monitoring leads according to the identification of the algorithm.
HR measurement range and accuracy:
Measurement range:
Adult: 10~300 bpm
Pediatric/Neonatal: 10~350bpm。
Resolution 1 bpm
Accuracy ±1% or ±1 bpm, whichever is greater
Detection ≥0.20mVp-p sensitivity
ST Display: Display 12-lead ST segment values at the same time and support ST graphic display.
ST measurement range and accuracy:
Measurement range: -2.0mV~2.0mV
Resolution ±0.01mV
Accuracy -0.8mV~0.8mV: ±0.02mV or ±10%, whichever is greater
Other: Unspecified
ST Update period: 10s
PVCs measurement: Indicates the number of PVC in the past minute, ranging from 0/min ~150/min.
QT analysis function:
Measurement range
QT: 200ms~700ms
QTc: 200ms~700ms
deltaQTc: -500ms~500ms
QT-HR:
Adult: 15bpm~150bpm
Pediatric / neonatal: 15bpm~180bpm
Resolution: QT: 1ms / QTc: 1ms
Accuracy QT: ±30ms
Sampling rate: 1000Hz (The time bias among every channels ≤100us)
Amplitude quantisation ≤1 uV/LSB
Arrhythmia Events list
No. Full Name Prompt information
1 Asystole = Asystole
2 Ventricular Fibrillation = Vent Fib/Tach
Ventricular Tachycardia
3 Ventricular Tachycardia = V-Tach
4 Ventricular Bradycardia = Vent Brady
5 Extreme Tachycardia = Extreme Tachy
6 Extreme Bradycardia = Extreme Brady
7 R on T = R on T
8 Tachycardia = Tachy
9 Bradycardia = Brady
10 Nonsustained Ventricular Tachycardia = Nonsustained V-Tach
11 Ventricular Rhythm = Vent Rhythm
|12 Pacer Not Captured = PNC
13 Pacer Not Pacing = PNP
14 Heartbeat Pause = Pause
15 Pauses/min High Pauses/min High
16 Run PVCs Run PVCs
17 Couplet Couplet
18 Ventricular Bigeminy = Bigeminy
19 Ventricular Trigeminy = Trigeminy
20 Frequent PVCs = Frequent PVCs
21 Premature ventricular = contraction PVC
22 Missed Beat = Missed Beat
23 Atrial Fibrillation = A-Fib
24 Atrial Fibrillation End = A-Fib End
25 ECG Noise = ECG Noise
26 Irregular Rhythm = Irregular Rhythm|
27 Irregular Rhythm End Irregular = Rhythm End
RESP
Measurement parameter: RR and respiration waveform
Source: RA-LA, RA-LL (default)
|Excitation waveform: 64 kHz
Error: ≤±10%
Excitation current ≤0.3mA RMS
Respiration Apnea Alarm: Fixed high priority alarm Adjustable delay time: 10~60s, error ±3s or ±10%
Cardiac interference alert: Fixed high priority alarm
RR measurement range and accuracy: Measurement range 0~150 rpm
Resolution 1 rpm
Accuracy: ±2 rpm or ±2%, whichever is greater.
Bandwidth: 0.2 Hz ~2.5Hz
Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s, ±10%
Baseline impedance range: 200~2500Ω (using defibrillator proof cable with resistance of 1kΩ)|
Measuring impedance range: 0.3Ω~3Ω
Gain: X-0.25, x~0.5, x~1, x~2, x~4