...

Valleylab FT10 Electrosurgical Unit

$21,400 + GST

FT10 Electrosurgical Unit

Valleylab by Medtronic

The FT10 is a world-leading ESU platform with bloodlines back to the FORCE FX and EX models.
A full range of accessories can be custom designed for your specific applications including a ligearture package.

The price is for a base. unit only – all accessories are optional.
Please allow 8 – 10 weeks for delivery.


SKU VALL-FT10 Categories , ,

Medtronic logovalleylab logo

Share
Description

Valleylab FT10 Electrosurgical Unit

The Valleylab FT10 ESU not only drives our full portfolio of energy-based devices, it makes them better than ever.

Improving the performance of all of our devices with faster sealing and cutting times – for example, LigaSure vessel sealing is up to 50% faster – the Valleylab FT10 energy platform delivers precise amounts of energy where and when it is needed most.

The Valleylab energy portfolio’s best-in-class advanced energy and electrosurgery devices are all powered by Smart generators that manage energy delivery with precision-enabling algorithms. The technology progress created with the Valleylab FT10 energy platform improves all surgical applications for the entire portfolio.

Next-generation changes

When used with the Valleylab FT10 energy platform:

Technology is improved with:
up to 50% faster-sealing times
lower jaw temperatures

Electrosurgical performance is more precise:
Auto bipolar has a faster activation time
Monopolar performance is improved

 

Valleylab FT10 Electrosurgical Unit Specifications

Output configuration:  Isolated output
Cooling:  Natural convection and fan
Display: 7 in. LCD touchscreen
Connector receptacles: LED illuminated connector readers on the LigaSure/Bipolar receptacle

Enclosure: Magnesium

Mounting
• Valleylab Universal Generator Cart (VLFTCRT)
• Operating-room boom systems
• Any stable, flat surface such as a table or cart top

Operating System: Linux™*

General Dimensions and Weight: Height 7.0 in. (17.78 cm) Width 14.5 in. (35.8 cm) Length 18.2 in. (46.2 cm) Weight 22.3 lb. (10.1 kg)

Environmental Parameters

Operation:
Ambient temperature range: 50 to 104 °F (10 to 40 °C)
Atmospheric pressure: 700 to 1060 millibars 500 to 1060 millibars
Relative humidity:  25% to 85% non-condensing

Transport and Storage:
Ambient temperature range:  14 to 140 °F (-10 to +60 °C)
Atmospheric pressure: 500 to 1060 millibars
Relative humidity: 30% to 75% non-condensing

1. The system can be stored for up to one year without performance degradation upon use.
2. If the generator is stored at a temperature outside the normal operating range of 10 to 40 °C, the system is ready for use after at least one hour at ambient temperature 20 °C ± 5 °C.

Input Power:

200–240 VAC Units
Line Ranges:
Line Voltage Full Regulation Range 180–264 VAC
Line Frequency 47–63 Hz
Max VA nominal line Voltage 950 VA
Max Mains Current: 6 A
Max Heat Dissipation: 180 W
Fuses 10 Fuses (2) –  5 mm x 20 mm 6.3 A, 250 V fast blow
Power Cord NEMA 3-prong hospital-grade connector

Additional information
Weight 20 kg