Description
Nellcor PM10N pulse oximeter
Nellcor_PM10N_Brochure
PM10N_OperatorsManual_en_PT00106388A00
The Nellcor™ Portable SpO2 Patient Monitoring System effectively monitors a broad range of patients across care areas and provides connectivity and analytic’s for robust data analysis.
Features and Benefits
- Small, lightweight, ergonomic handheld monitor for continuous and spot-check pulse oximetry monitoring
- Supports a robust data set, including real-time SpO2 and PR values, SatSeconds alarm management, pleth waveform, blip bar and tabular trend data
- Supports standard and home-care modes for use in hospital, hospital-type facilities, transport, mobile environments and home-care environments
- Sleep Study Mode, which enables the clinician to dim the LCD display and silence alarms to prevent disruption of patients’ sleep
- Nellcor™ advanced signal processing technology delivers reliable operation even in difficult monitoring conditions
- LoSat expanded accuracy range (60% to 100% SpO2) when used with Nellcor™ adhesive sensors with OxiMax™ technology
- Compatible with the complete line of Nellcor™ single-patient-use, reusable and specialty sensors with OxiMax™ technology
- 80 hours of data storage capability
- Supports wired data export to an external personal computer for data analysis and printing functions
- Can use 4 regular AA Batteries which will operate for 4-8 hours or use 4 lithium AA Batteries which operate for 20 hours.
- Available with Nellcor Analytics Software for Free!
Materials Included
- Portable SpO2 Patient Monitoring Unit
- 1 Adult Durasensor Reusable Sensor
- 1 Dura-Y Sensor
- 1 Reusable Pediatric Spot Check Clip
- 4 AA Lithium Batteries
- 1 User Manual
Nellcor PM10N pulse oximeter Specifications
Enclosure
Weight 274 g (0.604 lbs), including four batteries
Dimensions 70 mm W x 156 mm H x 32 mm D
Display
Screen size 88.9 mm (3.5 in), measured diagonally
Screen type TFT LCD, white LED backlight, viewing cone of 60º and optimal viewing distance of 1 meter
Resolution 320 x 480 pixels
Alarms
Categories: Patient status and system status
Priorities: Low, medium and high
Notification: Audible and visual
|Setting: Default, institutional and last setting
Alarm volume level: 49 to 89 dB Alarm system delay
Range and Accuracy
Range Values Measurement:
Ranges SpO2 saturation range: 1% to 100%
Pulse rate range: 20 to 250 beats per minute (bpm)
Perfusion range 0.03% to 20%
Display sweep speed 6.25 mm/sec
Measurement Accuracy
Saturation Adult: 70% to 100% ±2 digits
Saturation Adult and neonate low sat: 60 to 80% ±3 digits
Saturation Neonate: 70 to 100% ±2 digits
Low perfusion: 70 to 100% ±2 digits
Adult and neonate with motion: 70 to 100% ±3 digits
Pulse rate:
Adult and neonate: 20 to 250 bpm ±3 digits
Low perfusion 20 to 250 bpm ±3 digits
Adult and neonate with motion 20 to 250 bpm ±5 digits
Additional information
Weight | 5 kg |
---|