Biolight M800 Neonatal Pulse Oximeter
A great oximeter for your busy department, practice or home, rugged and reliable, hundreds sold in NZ and Australia over the years.
Kit Includes:
Main Unit – AA Battery Operation
BLT SpO2 – Neonatal wrap sensor
Shown with optional stand
Limited supply, more on order 3-5 week lead time.
Description
Biolight M800 Neonatal Pulse Oximeter
A fab little unit with hundreds of units sold in the AU market. This can be supercharged with Nellcor™ Oximetry and ECG
- Auto rotation Hi-Res 2.4″ colour OLED display: 320×240 pixels
- SPO2 and PR monitoring (ECG optional)
- Suitable for Adult, Pediatric & Neonatal patients
- Spot check and continuous monitoring modes
- Max 60,000 group short trend data store & review
Biolight M800 Neonatal Pulse Oximeter – Specifications:
Mainframe weight: <400g (full configuration, including the batteries)Mainframe size: 58.5mm(W)Å~123mm(H)Å~28mm(D)
Environmental Specifications:
Temperature: Operating: 5 ℃ to +40℃
Storage: -20 ℃ to +55℃
Atmospheric: Operating: 700hPa to 1060hPa
pressure Storage: 500hPa to 1060hPa
Humidity: Operating: 15% to 85%(non-condensing)
Storage: 10% to 93%(non condensing)
Charging Specifications: (optional)
AC-DC Adapter: Input: 100 ~240VAC, 50/60Hz , 0.5A
Output: 5V, 1.5A
Battery Specification:
Type: 1.5V, AA alkaline battery
Capacity: 2000mAh
Quantity: 3
Run time: >14 hours
Shutdown delay: 10 min(After the first “low battery” alarm)
Display:
Size: 2.4 inch
Resolution: 320*240 pixels
Visual Alarm Indicator:
Alarm indicating 1amp:(Yellow/Red)
Battery charging indicating lamp: 1 (orange)
When charged, it lights orange.
When fully charged or not charged, it doesn’t light.
Data Storage:
Displaying way: Trend tabular
Trend interval: 10 seconds to 30 minutes
Trend parameter: HR, SpO2
Trend data:
Spot-check: ID from 1 to 99, 300 groups per ID.
Continuous: ID is 0, 30000 groups can be stored.
BLT SpO2:
Measurement range: 0~100%
Resolution: 1%
Accuracy: 70~100% :± 2%
0% to 69%: unspecified
Refreshing rate: <13 seconds
Pitch Tone: Yes
Nellcor SpO2 (Optional):
Measurement range : 25 bpm ~250 bpm
Resolution: 1 bpm
Accuracy: ±1% or ±1 bpm, whichever is the greater
Refreshing rate: <13 seconds
Pitch Tone: Yes
PR:
Measurement range: 20 bpm ~300 bpm
Resolution: 1 bpm
Accuracy: 20 bpm ~250 bpm: ± 3 bpm
251 bpm~300 bpm: unspecified
Refreshing rate: 7 seconds
ECG Specifications: (optional)
Lead type: Standard: 3-lead (RA, LA, LL or R, L, F)
Gain: 2.5mm/mV(Å~0.25), 5mm/mV(Å~0.5), 10mm/mV(Å~1)
Input impedance: ≥5.0MΩ
Input current: <0.1 uA
Baseline recovery: ≤ 3 s
ECG signal input range: -6.0mV to+6.0mV
CMRR: ≥90dB
Leakage Current: <10 uA
Frequency response: 0.5Hz – 40Hz
Standardizing signal: 1mV ± 5%
Defibrillator-proof
Electrode off indicating: Yes
HR range: 10 bpm ~300 bpm
HR Resolution: 1 bpm
HR Accuracy: ±1% or ±1 bpm, whichever is the greater
Neonatal sp02
Additional information
Weight | 5 kg |
---|